6
•
Operation
Integral Fraction Collector FT (VF-F20-A) Operating Manual
Page 133
No.
Description
No.
Description
1
Lock pin
3
Pusher
2
Pusher cover
4
Needle
6.8.1
Adjust to AboveVial Pusher Position
Preparations
1. Open the left front door
.
2. Press the service button on keypad to move the needle unit into
service position.
NOTICE
To avoid personal injury, do not put hands into the
compartment when the needle unit is moving.
Follow these steps
1. Raise the lock pin upwards by hand. The pusher will be moved up
accordingly.
Figure 52: Moving the pusher up by hand
No.
Description
No.
Description
1
Lock tab
2
U-shape slot
Содержание VF-F20-A
Страница 4: ...Contacting Us Page 4 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 17: ...1 Using this Manual Integral Fraction Collector FT VF F20 A Operating Manual Page 17 ...
Страница 31: ...2 Safety Integral Fraction Collector FT VF F20 A Operating Manual Page 31 ...
Страница 48: ...3 Device Overview Page 48 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 215: ...8 Troubleshooting Integral Fraction Collector FT VF F20 A Operating Manual Page 215 ...
Страница 221: ...9 Specifications Integral Fraction Collector FT VF F20 A Operating Manual Page 221 ...
Страница 231: ...10 Accessories Consumables and Replacement Parts Integral Fraction Collector FT VF F20 A Operating Manual Page 231 ...
Страница 236: ...12 Index Page 236 Integral Fraction Collector FT VF F20 A Operating Manual 12 Index ...
Страница 240: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...