6
•
Operation
Integral Fraction Collector FT (VF-F20-A) Operating Manual
Page 107
•
Monitoring the detector signal and checking whether the detector
signal is stable so that the drift and signal noise are within
reasonable limits for the application.
•
Performing an autozero of the detector baseline.
TIP
The Chromeleon software supports procedures for automatically
starting a chromatography system in the software (
Smart Startup
). The
startup procedure includes the operations for system equilibration. For
details, refer to the
Chromeleon Help
.
6.5.1
Thermostatting the Fraction Collector Compartment
To use thermostatting, define the following parameters:
•
Temperature control
To thermostat the fraction collector compartment, enable the
temperature control (
Temperature Control = On
).
•
Target temperature
Define the temperature setpoint to which the fraction collector
compartment is to be cooled or heated (
Temperature Nominal
).
Observe the following:
Ensure that the thermostatting temperature is suitable for your
samples and within the specified temperature range of the
device.
When you enter a temperature while the temperature control is
disabled, temperature control will be enabled.
If the device is turned off and on again, the temperature control
will be disabled.
6.5.2
Loading the Carousel
The carousel separates into four color-coded segments: red (
R
),
green (
G
), blue (
B
) and yellow (
Y
).
For further information on the sample compartment in general, the
carousel as well as the rack type identification, see
(
Page 43)
.
Содержание VF-F20-A
Страница 4: ...Contacting Us Page 4 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 17: ...1 Using this Manual Integral Fraction Collector FT VF F20 A Operating Manual Page 17 ...
Страница 31: ...2 Safety Integral Fraction Collector FT VF F20 A Operating Manual Page 31 ...
Страница 48: ...3 Device Overview Page 48 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 215: ...8 Troubleshooting Integral Fraction Collector FT VF F20 A Operating Manual Page 215 ...
Страница 221: ...9 Specifications Integral Fraction Collector FT VF F20 A Operating Manual Page 221 ...
Страница 231: ...10 Accessories Consumables and Replacement Parts Integral Fraction Collector FT VF F20 A Operating Manual Page 231 ...
Страница 236: ...12 Index Page 236 Integral Fraction Collector FT VF F20 A Operating Manual 12 Index ...
Страница 240: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...