3 • Operation and Maintenance
Doc. 065259-05 10/12
39
•
To interrupt the autosampler when a sequence is
not
running, click
Stop
on the Dionex AS-DV ePanel in Chromeleon 7 or the Dionex
AS-DV Control panel in Chromeleon 6.8.
NOTE After the Dionex AS-DV stops its current
operation, the carousel returns to the home
position (position 1). Do not reach into the
autosampler while the carousel is moving!
3.8.2
Specifying a Bleed
A bleed delivers excess sample (and any trapped air) from the vial to
waste, before delivering the injection volume to the sample loop or
concentrator column. Excess sample is any sample above the fill line on
the vial (see
You can use either the injection valve or an optional high-pressure valve
configured as an auxiliary valve to perform the bleed. For details about
plumbing a high-pressure valve as a bleed valve, see
.
Enabling a Bleed
1. If you are creating a new instrument method or program, go to the
Sample Preparation Options page of the Instrument Method Wizard
or Program Wizard (see
If you are editing an existing instrument method or program, go to the
Sample Preparation page of the Instrument Method Editor or Program
Editor.
2. If you are using
Basic Edit Mode
: In the
Bleed
list, select the valve
position that will direct the bleed to waste. For the typical injection
valve plumbing, select
InjectPosition
). For the
typical auxiliary valve plumbing, select
AuxValve.A
Содержание Dionex AS-DV
Страница 1: ...Dionex AS DV Autosampler Operator s Manual Document No 065259 Revision 05 October 2012 ...
Страница 8: ...Dionex AS DV Operator s Manual vi Doc 065259 05 10 12 ...
Страница 30: ...Dionex AS DV Operator s Manual 22 Doc 065259 05 10 12 ...
Страница 126: ...Dionex AS DV Operator s Manual 116 Doc 065259 05 10 12 954745 Fuse IEC 127 rated 3 15 A 1 Part Number Item Quantity ...
Страница 132: ...Dionex AS DV Operator s Manual Index 6 Doc 065259 05 10 12 ...