2 • Description
Doc. 065259-05 10/12
15
2.2
Dionex AS-DV
Operating Principles
2.2.1
Sampling Operating Principles
The sampling head tower (see
) consists of a motor-driven
needle, with a sampling tip on the end, that moves down to engage the top
of the vial cap.
As the needle moves downward, the cap is gradually pushed into the vial
(see
), forcing the liquid through the filter (if present) and into
the sample out line. The moving cap acts as a piston to force the liquid
from the vial. This allows samples to be loaded without the use of an
external sampling pump. The concave bottom of the vial cap traps air, so
that it can be expelled with any excess sample above the fill line. In
addition, the shape exactly matches the convex bottom of the sample vial;
when the cap is pushed completely into the vial, virtually all of the sample
is expelled.
The socket for the sampling tip is identical on both sizes of caps to allow
mating with the tip of the sampler head. The sampling head speed
(delivery speed) is set in Chromeleon (see
Figure 2-7. PolyVial, Cap, and Sampling Tip Features
Sampling Tip
Sample
Cap
20 m Filter (Optional)
P
Trapped Air and Excess Sample
(Expelled before sample delivery
if Bleed is enabled)
Convex vial bottom mates
with concave cap to ensure
delivery of sample
complete
Sample Out Line
Fill Line
Needle
Содержание Dionex AS-DV
Страница 1: ...Dionex AS DV Autosampler Operator s Manual Document No 065259 Revision 05 October 2012 ...
Страница 8: ...Dionex AS DV Operator s Manual vi Doc 065259 05 10 12 ...
Страница 30: ...Dionex AS DV Operator s Manual 22 Doc 065259 05 10 12 ...
Страница 126: ...Dionex AS DV Operator s Manual 116 Doc 065259 05 10 12 954745 Fuse IEC 127 rated 3 15 A 1 Part Number Item Quantity ...
Страница 132: ...Dionex AS DV Operator s Manual Index 6 Doc 065259 05 10 12 ...