Dionex AS-DV Operator’s Manual
32
Doc. 065259-05 10/12
Estimator
to estimate the volume of sample the Dionex AS-DV
will deliver to the loop or concentrator column. The estimate is
provided for informational purposes only. In the
Loop Volume
box, enter the volume of the loop installed on the injection valve.
5. After specifying the Dionex AS-DV options, click
Next >
and
complete the remaining wizard pages.
3.5.2
Dionex AS-DV General Parameters
The following general parameters are included on the Sampler
Preparation Options page.
Delivery Speed
: Determines the speed at which the sample is delivered to
the sample loop or concentrator column. The following table shows the
recommended delivery speeds:
Delay Volume
: Specifies the volume from the tip of the sampling needle
to the injection valve. The range is
0 to 500 µL. The default value of
125 µL is appropriate for the 1.4 m (56 in) length of sample out line
provided with the Dionex AS-DV. You can set the
Delay Volume
in the
Chromeleon 7 Instrument Configuration Manager or the Chromeleon 6.8
Server Configuration program (see
Cycle Time:
To set a uniform time between injections, select this check
box and enter the time in minutes. The time between injections is never
shorter than the specified cycle time. However, if the previous instrument
method or program is still running when the cycle is over, cycle time will
be ignored. Clear the
Cycle Time
check box to let the commands entered
in the instrument method or program determine the time between
injections.
Flush Factor
: Chromeleon uses this parameter in the formula that
determines the volume of sample delivered from the vial. The formula is
shown in the
Total Volume Delivered Estimator
.
Vial Size
Type
Speed (mL/min)
Small (0.5 mL)
Loop
1.6
Concentrator
0.4
Large (5.0 mL)
Loop
4.0
Concentrator
1.0
Содержание Dionex AS-DV
Страница 1: ...Dionex AS DV Autosampler Operator s Manual Document No 065259 Revision 05 October 2012 ...
Страница 8: ...Dionex AS DV Operator s Manual vi Doc 065259 05 10 12 ...
Страница 30: ...Dionex AS DV Operator s Manual 22 Doc 065259 05 10 12 ...
Страница 126: ...Dionex AS DV Operator s Manual 116 Doc 065259 05 10 12 954745 Fuse IEC 127 rated 3 15 A 1 Part Number Item Quantity ...
Страница 132: ...Dionex AS DV Operator s Manual Index 6 Doc 065259 05 10 12 ...