2 • Description
Doc. 065259-05 10/12
17
show the flow schematics for the 6-port and 10-
port valves.
Operation as an Application Valve
The optional valve can be configured for various application functions,
including sample injection, concentrator column loading, or matrix
elimination.
Plumbing configurations for the valves depend on the components to be
connected and the application to be run. Refer to the appropriate Thermo
Scientific Dionex Application Note for more information.
Operation as a Bleed Valve
When the optional valve is configured as a bleed valve (see
and
), the valve directs excess sample and any air trapped in
the sample vial to waste. Excess sample is any sample above the fill line
on the vial (see
). After the initial bleed, the valve position is
switched and the sample is sent directly to the injection valve in the IC
system.
Figure 2-8. Optional 6-Port HIgh-Pressure Valve Flow Schematics
Figure 2-9. Optional 10-Port High-Pressure Valve Flow Schematics
B Position
4
1
5
6
2
3
4
1
5
6
2
3
1
3
5
7
9
2
4
6
8
10
1
3
5
7
9
2
4
6
8
10
B Position
Содержание Dionex AS-DV
Страница 1: ...Dionex AS DV Autosampler Operator s Manual Document No 065259 Revision 05 October 2012 ...
Страница 8: ...Dionex AS DV Operator s Manual vi Doc 065259 05 10 12 ...
Страница 30: ...Dionex AS DV Operator s Manual 22 Doc 065259 05 10 12 ...
Страница 126: ...Dionex AS DV Operator s Manual 116 Doc 065259 05 10 12 954745 Fuse IEC 127 rated 3 15 A 1 Part Number Item Quantity ...
Страница 132: ...Dionex AS DV Operator s Manual Index 6 Doc 065259 05 10 12 ...