Doc. 065259-05 10/12
69
5 • Service
This chapter describes Dionex AS-DV service and repair procedures that users
may perform. Procedures not included here, including electronics-related repair
procedures, must be performed by Thermo Fisher Scientific personnel. For
assistance, contact Technical Support for Dionex products. In the U.S. and
Canada, call 1-800-346-6390. Outside the U.S. and Canada, call the nearest
Thermo Fisher Scientific office.
Before replacing any part, refer to the troubleshooting information in
to
correctly identify the cause of the problem.
5.1
Replacing Tubing and Fittings
The Dionex AS-DV is plumbed with the needle and tubing assembly listed below:
Dionex AS-DVtubing connections use high-pressure 10-32 fitting bolts
(P/N 074449) and high-pressure double-cone ferrules (P/N 074373).
To tighten a high-pressure fitting bolt and ferrule:
1. Use your fingers to tighten the fitting bolt as tight as you can.
2. Then, use a wrench to tighten the fitting an additional three-quarter turn
(270 degrees).
3. If leaks occur, replace the fitting bolt, ferrule, and tubing.
Substituting non-Dionex/Thermo Scientific parts may impair a module’s
performance, thereby voiding the product warranty. Refer to the
warranty statement in the Dionex Terms and Conditions for more
information.
Tubing Assembly
Used for
Dionex AS-DV Sampling Tip Kit
(P/N 071575)
Connection from sampling head to
injection valve
Содержание Dionex AS-DV
Страница 1: ...Dionex AS DV Autosampler Operator s Manual Document No 065259 Revision 05 October 2012 ...
Страница 8: ...Dionex AS DV Operator s Manual vi Doc 065259 05 10 12 ...
Страница 30: ...Dionex AS DV Operator s Manual 22 Doc 065259 05 10 12 ...
Страница 126: ...Dionex AS DV Operator s Manual 116 Doc 065259 05 10 12 954745 Fuse IEC 127 rated 3 15 A 1 Part Number Item Quantity ...
Страница 132: ...Dionex AS DV Operator s Manual Index 6 Doc 065259 05 10 12 ...