1 • Introduction to the Dionex AS-DV
Doc. 065259-05 10/12
5
1.3.2
Safety Symbols
The symbols below appear on the Dionex AS-DV or on Dionex AS-DV
labels.
1.4
Deionized Water Requirements for IC
For all procedures requiring deionized water, use ASTM Type I (18 megohm-cm)
filtered and deionized water that meets the specifications listed in
Alternating current
Primary protective conductor terminal
Secondary protective conductor terminal
Power supply is on
Power supply is off
Indicates a potential hazard. Refer to this operator’s
manual for an explanation of the hazard and how to
proceed.
Contaminant
Specification
Ions–Resistivity
>18.0 (megohm-cm)
Organics–TOC
<10 ppb
Iron/Transition Metals*
<1 ppb
Pyrogens
<0.03 (Eu/mL)
Particulates > 0.2 µm
<1 (units/mL)
Colloids–Silica
<10 ppb
Bacteria
<1 (cfu/mL)
* Iron/transition metal content not specified for ASTM Type I Water
Table 1-1. ASTM Filtered, Type I Deionized Water Specifications