3 • Operation and Maintenance
Doc. 065259-05 10/12
51
For example, to take five injections from a filled 5.0 mL vial, select
1000
P
L. In the instrument method Script or program Commands
view, a
DeliverSample Volume=1000
command appears (see
Notes:
•
The wizard adds the
DelayVolume
and
FlushFactor
commands to
the instrument method or program automatically. However, they are
ignored in this example. If a second
DeliverSample
command
(without a specified volume) were included, the
DelayVolume
and
FlushFactor
commands would be used to calculate the volume for
the second
DeliverSample
command.
•
In the Script or Commands view, you can enter the
Volume
for the
DeliverSample
command in 1
P
L increments. For example, to take
three injections from a 5000
P
L vial, enter the following command:
DeliverSample
Volume=1666
.
Figure 3-16. Chromeleon 6.8 Commands View Example: Deliver Multiple Injections
Per Vial by Specifying the Volume