5. Remove the 6-port or 10-port high-pressure valve:
a. Use a 3-mm Allen wrench to remove the two screws holding the valve
assembly in place (see
b. Slide the valve assembly partway out of the chassis, press the locking
latch on the cable (see
), and disconnect the cable.
c. Remove the valve assembly.
6. Follow the instructions provided in the Rebuild Kit to replace the rotor seal,
isolation seal, and stator face.
Figure 5-13. Removing the Valve Assembly
Figure 5-14. Disconnecting the Valve Cable
Remove these
2 screws
Locking Latch