Dionex AS-DV Operator’s Manual
102
Doc. 065259-05 10/12
B.8.4
Verifying Chromeleon Communication
To verify that Chromeleon is communicating with the Dionex AS-DV:
1. On the Dionex AS-DV ePanel (in Chromeleon 7) or Control panel (in
Chromeleon 6.8), under
Settings
, enter a
Vial Position
.
2. Under
Commands
, click
Move to Vial
. The carousel should move to
the entered position.
3. If an optional valve is installed: In Chromeleon 7, click the
Inject
Position
and then the
Load Position
button. In Chromeleon 6.8, click
the
Position A
and then the
Position B
button. You should hear the
valve actuate as it changes position.
4. If the carousel or valve do not respond, check the Chromeleon audit
trail for error messages. For troubleshooting information, refer to
in this manual or to the Chromeleon Help.
B.8.5
Setting Up the Dionex AS-DV for Simultaneous Injections
The simultaneous injection feature allows delivery of sample to two
independent ion chromatography (IC) systems with one Dionex AS-DV
and a computer running Chromeleon. The Dionex AS-DV and AS40
Simultaneous Injection Kit for ICS Systems (P/N 060258) provides the
necessary parts and instructions for setting up this feature.
Содержание Dionex AS-DV
Страница 1: ...Dionex AS DV Autosampler Operator s Manual Document No 065259 Revision 05 October 2012 ...
Страница 8: ...Dionex AS DV Operator s Manual vi Doc 065259 05 10 12 ...
Страница 30: ...Dionex AS DV Operator s Manual 22 Doc 065259 05 10 12 ...
Страница 126: ...Dionex AS DV Operator s Manual 116 Doc 065259 05 10 12 954745 Fuse IEC 127 rated 3 15 A 1 Part Number Item Quantity ...
Страница 132: ...Dionex AS DV Operator s Manual Index 6 Doc 065259 05 10 12 ...