B • Installation
Doc. 065259-05 10/12
89
Two connection options are available:
•
Connect the Dionex AS-DV sample out line directly to the injection valve in
the IC system (see
•
Connect the Dionex AS-DV sample out line to an optional 6-port or 10-port
high-pressure valve installed in the Dionex AS-DV (see
B.3.1
Connecting the Sample Out Line Directly to an IC System
1. Route the sample out line through the side slot in the Dionex AS-DV
chassis. Leave a little slack in the tubing.
2. Locate a 10-32 fitting bolt (P/N 074449) and
double-cone ferrule (P/N 074373) in the Dionex
AS-DV Ship Kit (for 0.5 mL vials, P/N AAA-
071115; for 5.0 mL vials, P/N 068887).
3. Install the fitting and ferrule onto the end of the
Dionex AS-DV sample out line. See
for installation and tightening requirements.
Figure B-2. Routing the Dionex AS-DV Sample Out Line to an IC System
Route sample
out line through
side slot to IC
system