Dionex AS-DV Operator’s Manual
84
Doc. 065259-05 10/12
5.7
Replacing the Main Power Fuses
1. Turn off the power switch on the Dionex AS-DV rear panel (see
2. Disconnect the main power cord from both its source and from the Dionex
AS-DV rear panel.
3. The fuse drawer is located below the main power switch (see
4. A recessed lock is located on each side of
the fuse drawer. Use a small screwdriver
(or your fingernail) to push each lock
toward the center to release it. When both
locks are released, pull the fuse drawer
straight out of its compartment.
5. Replace the two fuses with new 3.15 A
IEC 127 fast-blow fuses (P/N 954745).
Thermo Scientific recommends always
replacing
both
fuses.
Figure 5-18. Dionex AS-DV Fuse Drawer
6. Insert the fuse drawer into its compartment. Apply sufficient pressure evenly
against the drawer to engage the two locks. The drawer will be flush against
the panel when both locks are engaged.
7. Reconnect the main power cord and turn on the power.
HIGH VOLTAGE—Disconnect the main power cord from its source and
also from the rear panel of the Dionex AS-DV.
HAUTE TENSION—Débranchez le cordon d'alimentation principal de
sa source et du panneau arrière du Dionex AS-DV.
HOCHSPANNUNG—Ziehen Sie das Netzkabel aus der Steckdose und
der Netzbuchse auf der Rückseite des Dionex AS-DV.
Locks
Fuse
Drawer
Power
Switch
Содержание Dionex AS-DV
Страница 1: ...Dionex AS DV Autosampler Operator s Manual Document No 065259 Revision 05 October 2012 ...
Страница 8: ...Dionex AS DV Operator s Manual vi Doc 065259 05 10 12 ...
Страница 30: ...Dionex AS DV Operator s Manual 22 Doc 065259 05 10 12 ...
Страница 126: ...Dionex AS DV Operator s Manual 116 Doc 065259 05 10 12 954745 Fuse IEC 127 rated 3 15 A 1 Part Number Item Quantity ...
Страница 132: ...Dionex AS DV Operator s Manual Index 6 Doc 065259 05 10 12 ...