Dionex AS-DV Operator’s Manual
62
Doc. 065259-05 10/12
This error occurs when a command is sent while the autosampler is
performing another function.
To troubleshoot:
Wait for one function to finish before issuing a command to begin another.
This error occurs if a controlling device sends a signal to the TTL input while
the Dionex AS-DV is already performing a command.
To troubleshoot:
1. After the other command is completed, send the signal to the TTL input
again. For details about the time required to complete the sample load
cycle, refer to
.
2. If the problem persists, contact Technical Support for Dionex products for
assistance.
This error occurs if the optional high-pressure valve is not connected properly
or if the valve fails to switch position within 1 second of being toggled.
To troubleshoot:
1. Check the valve connections:
a. Start the Chromeleon 7 Instrument Configuration Manager or the
Chromeleon 6.8
Server Configuration program:
•
For Chromeleon 7, click
Start
on the Windows taskbar and click
All Programs > Chromeleon 7 > Instrument Configuration
Manager
.
•
For Chromeleon 6.8, click
Start
on the Windows taskbar and
click
All Programs > Chromeleon > Server Configuration
.
This command is not allowed at this stage because the
autosampler is busy.
TTL triggered while another command is in progress.
Valve error.
Содержание Dionex AS-DV
Страница 1: ...Dionex AS DV Autosampler Operator s Manual Document No 065259 Revision 05 October 2012 ...
Страница 8: ...Dionex AS DV Operator s Manual vi Doc 065259 05 10 12 ...
Страница 30: ...Dionex AS DV Operator s Manual 22 Doc 065259 05 10 12 ...
Страница 126: ...Dionex AS DV Operator s Manual 116 Doc 065259 05 10 12 954745 Fuse IEC 127 rated 3 15 A 1 Part Number Item Quantity ...
Страница 132: ...Dionex AS DV Operator s Manual Index 6 Doc 065259 05 10 12 ...