4 • Troubleshooting
Doc. 065259-05 10/12
61
This error occurs when the needle is jammed or there is excessive
backpressure.
To troubleshoot:
1. Press the
Carousel Release
button to release the carousel and raise the
needle. If the needle does not raise automatically, manually raise the
needle (see
2. Check for blockages.
3. Check for excessive backpressure (see
4. If the problem persists, contact Technical Support for Dionex products for
assistance.
This error occurs when there is insufficient liquid in the vial to deliver the
requested volume.
To troubleshoot:
Add more sample to the vial or select a smaller volume in Chromeleon (see
This error occurs when the carousel is blocked or the sample head is jammed.
To troubleshoot:
1. Remove the carousel (see
) and examine for blockages.
2. If the problem persists, replace the sampling tip (see
3. If the problem continues, contact Technical Support for Dionex products
for assistance.
Needle is jammed.
Not enough volume to deliver.
Sample needle failed to reach destination.
Содержание Dionex AS-DV
Страница 1: ...Dionex AS DV Autosampler Operator s Manual Document No 065259 Revision 05 October 2012 ...
Страница 8: ...Dionex AS DV Operator s Manual vi Doc 065259 05 10 12 ...
Страница 30: ...Dionex AS DV Operator s Manual 22 Doc 065259 05 10 12 ...
Страница 126: ...Dionex AS DV Operator s Manual 116 Doc 065259 05 10 12 954745 Fuse IEC 127 rated 3 15 A 1 Part Number Item Quantity ...
Страница 132: ...Dionex AS DV Operator s Manual Index 6 Doc 065259 05 10 12 ...