Dionex AS-DV Operator’s Manual
70
Doc. 065259-05 10/12
5.2
Removing the Carousel
The following tools are needed to complete this procedure: #1 Phillips
screwdriver, 3-mm Allen wrench, and 2.5-mm Allen wrench.
1. Press the
Carousel Release
button to raise the sampling tip to its fully up
position.
2. Turn off the Dionex AS-DVpower and disconnect the power cord. Raise the
top cover.
3. Remove the Dionex AS-DV front cover by gripping the indentations at each
side and pulling the cover straight off toward you.
4. Use a #1 Phillips screwdriver to remove the three flathead screws from the
sampling head tower cover (see
). Lift up the cover to remove it.
Figure 5-1. Removing the Sampling Head Tower Cover
Remove 3 screws