4. Connect the Dionex AS-DV
sample out line to the
sample in port on the IC
system injection valve (see
5. Replace the front panel and
close the top cover on the
Dionex AS-DV.
Figure B-3. IC System Injection Valve
Go on to
B.3.2
Connecting the Sample Out Line to the Optional High-
Pressure Valve
Installing the Optional Valve
1. Verify that the power to the Dionex AS-DV is off.
2. Use a 3-mm Allen wrench to
remove the two screws
holding the valve cover plate
to the chassis. Save the screws.
The Dionex AS-DV power must be turned off before you install a valve.
If the power is on when you connect or disconnect the valve cable, the
Dionex AS-DV will be damaged.
Dionex AS-
DV Sample
Out Line
Remove these
2 screws