Dionex AS-DV Operator’s Manual
60
Doc. 065259-05 10/12
This error occurs if you press the
Carousel Release
button while a sample is
being delivered.
To troubleshoot:
Wait for a function to finish before trying to release the carousel.
This error occurs if you send a command while the carousel or needle is
performing a function.
To troubleshoot:
Wait for one function to finish before issuing a command to begin another.
This error occurs if there is not enough liquid in the vials in the rinse vial
range or in the specified vial (if a single vial is specified for the rinse).
To troubleshoot:
1. Press the
Carousel Release
button to clear the error and reset the system.
2. Install filled vials in the rinse vial range or in the specified position.
3. Begin the sampling process again.
This error occurs when the needle is jammed or there is excessive
backpressure. The motor speed is reduced and the Dionex AS-DV attempts to
continue delivery. If the problem persists, the run is stopped, and the “Needle
is jammed” error message is displayed. For troubleshooting, refer to the
“Needle is jammed” section below.
The Carousel Engage switch is inhibited during deliver.
Carousel or needle is busy.
More rinse volume than remaining has been requested.
The motor speed is reduced to
¼
of the previous one
because of excessive load.
Содержание Dionex AS-DV
Страница 1: ...Dionex AS DV Autosampler Operator s Manual Document No 065259 Revision 05 October 2012 ...
Страница 8: ...Dionex AS DV Operator s Manual vi Doc 065259 05 10 12 ...
Страница 30: ...Dionex AS DV Operator s Manual 22 Doc 065259 05 10 12 ...
Страница 126: ...Dionex AS DV Operator s Manual 116 Doc 065259 05 10 12 954745 Fuse IEC 127 rated 3 15 A 1 Part Number Item Quantity ...
Страница 132: ...Dionex AS DV Operator s Manual Index 6 Doc 065259 05 10 12 ...