Dionex AS-DV Operator’s Manual
108
Doc. 065259-05 10/12
C.2 Connecting to the Relay Output and TTL Input
1. Locate the twisted pair of
wires (P/N 043598) and 4-
position connector plug
(P/N 923852) (see
) in the Dionex
AS-DV Ship Kit (for 0.5 mL
vials, P/N AAA-071115; for
5.0 mL vials, P/N 068887).
Figure C-3. 4-Position Relay TTL Connector Plug
2. To connect to
RELAY OUT
, connect an active wire (red) to position 1 and a
ground wire (black) to position 2. To connect to
TTL IN
, connect an active wire
(red) to position 3 and a ground wire (black) to position 4.
To attach a wire to the plug, strip the end of the wire (if it is not already),
insert the wire into the plug, and use a screwdriver to tighten the locking
screw.
3. Plug the connector into the 4-pin connector on the Dionex AS-DV rear panel
4. Connect the wires from the Dionex AS-DV connector plug to the appropriate
connector pins on the controlling device. Additional connector plugs are
provided with other Dionex products.
NOTE Check the polarity of each connection. Connect signal
wires to signal (+) pins and ground wires to ground (-)
pins. If you reverse the wires when connecting the relay
output, the relay will not work correctly.
5. If you connected
TTL IN
, verify that the correct input control mode is selected
When attaching wires to the connector plug, be careful not to allow
stray strands of wire to short to an adjoining position on the
connector.
Position 1
Locking Screws
Position 4
Содержание Dionex AS-DV
Страница 1: ...Dionex AS DV Autosampler Operator s Manual Document No 065259 Revision 05 October 2012 ...
Страница 8: ...Dionex AS DV Operator s Manual vi Doc 065259 05 10 12 ...
Страница 30: ...Dionex AS DV Operator s Manual 22 Doc 065259 05 10 12 ...
Страница 126: ...Dionex AS DV Operator s Manual 116 Doc 065259 05 10 12 954745 Fuse IEC 127 rated 3 15 A 1 Part Number Item Quantity ...
Страница 132: ...Dionex AS DV Operator s Manual Index 6 Doc 065259 05 10 12 ...