Setting waveform display parameters
Use waveform display controls to set the display mode, persistence, style, and intensity display parameters, and graticule style
and intensity.
Use the following topics for more information on setting display parameters.
■
Set waveform display mode (Stacked or Overlay)
on page 129
■
Set the Waveform Interpolation mode
on page 129
■
Set the waveform persistence, style, and intensity
■
Set the graticule style and intensity
on page 130
Set waveform display mode (Stacked or Overlay)
Use this procedure to change the waveform display mode (Stacked or Overlay).
1. Double-tap on an open graticule area to open the Waveform View configuration menu.
2. Tap the Display Mode buttons to toggle between Overlay and Stacked modes.
In stacked display mode, each waveform is stacked vertically in separate graticule slices. This is the default display mode.
In overlay display mode, all waveforms are displayed in a single graticule (traditional waveform view).
Set the Waveform Interpolation mode
Use this procedure to set the waveform interpolation mode, which sets how waveform data points are calculated between record
data points.
1. Double-tap on an open graticule area to open the Waveform View configuration menu.
2. Tap the buttons under Interpolation to select either Sin(x)/x or Linear.
Sin(x)/x calculates record points along a curve between the actual acquired samples. This form of interpolation is useful
when acquiring rounded waveforms such as sine waves. It is good for general-purpose uses but may introduce overshoot or
undershoot in signals with fast rise times. This interpolation is also useful for looking at high-frequency signals, especially
where the frequency components are just below the Nyquist frequency.
Linear calculates record points between actual acquired samples using a straight-line fit. This interpolation is useful for
measuring waveforms with fast rise times, such as pulse trains.
MSO54, MSO56, MSO58, MSO58LP, MSO64 Help
129
Содержание 6 series
Страница 24: ...Product documents and support 4 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 42: ...Options 22 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 54: ...Install your instrument 34 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 63: ...9 FlexChannel probe connectors Getting acquainted with your instrument MSO54 MSO56 MSO58 MSO58LP MSO64 Help 43 ...
Страница 84: ...Getting acquainted with your instrument 64 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 102: ...Configure the instrument 82 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 148: ...Advanced triggering 128 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 154: ...Zooming on waveforms 134 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 438: ...Waveform acquisition concepts 418 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 448: ...Waveform display concepts 428 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 483: ...Where σ is the Phase Angle Degree Measurement algorithms MSO54 MSO56 MSO58 MSO58LP MSO64 Help 463 ...
Страница 518: ...Index 498 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...