Set Trigger Holdoff
Trigger Holdoff sets the time, after triggering on an event, that the instrument waits before detecting the same trigger event to
start the next acquisition.
Setting the correct holdoff time is important to get a stable trigger. The longer holdoff time for the top waveform causes unstable
triggering. The shorter holdoff set for the bottom waveform only triggers on the first pulse in the burst to remedy the unstable
trigger.
Prerequisite: You have set up a trigger event for a signal.
1. Double-tap the Trigger badge on the Settings bar to open the Trigger configuration menu.
2. Tap the Mode & Holdoff panel.
3. To set a specific holdoff time, tap Time, then tap Holdoff Time and use the assigned multipurpose knob to specify a holdoff
time. Or double-tap the field and use the virtual keypad to enter a holdoff time.
4. To delay the trigger a random amount of time between triggers, tap Random.
Trigger on sequential events (A and B triggers)
Use the A and B Trigger Events to trigger on a second event after a first event occurs.
Set up sequential triggering
1. Double-tap the Trigger badge to open the Trigger configuration menu.
2. Tap Trigger Type and select Sequence.
3. Set up the A Trigger Event:
a. Tap the A Trigger Event button to open the A Trigger Event menu.
b. Tap Trigger Type and select a trigger type from the drop-down list. This example uses Edge.
c. Tap Source and select the A event trigger source.
d. If displayed, tap Coupling and select the trigger coupling.
e. Tap Level and set the desired trigger level using the multipurpose knob. Or double-tap the field to set the value using
the virtual keypad.
f.
Tap a Slope button to select the slope of the signal on which to trigger (Rise, Fall, Either).
g. Tap anywhere outside the A Trigger Event menu to close it.
4. Set up the B Trigger Event:
a. Tap the B Trigger Event button in the main Trigger configuration menu to open the B Trigger Event menu.
Advanced triggering
MSO54, MSO56, MSO58, MSO58LP, MSO64 Help
119
Содержание 6 series
Страница 24: ...Product documents and support 4 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 42: ...Options 22 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 54: ...Install your instrument 34 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 63: ...9 FlexChannel probe connectors Getting acquainted with your instrument MSO54 MSO56 MSO58 MSO58LP MSO64 Help 43 ...
Страница 84: ...Getting acquainted with your instrument 64 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 102: ...Configure the instrument 82 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 148: ...Advanced triggering 128 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 154: ...Zooming on waveforms 134 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 438: ...Waveform acquisition concepts 418 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 448: ...Waveform display concepts 428 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 483: ...Where σ is the Phase Angle Degree Measurement algorithms MSO54 MSO56 MSO58 MSO58LP MSO64 Help 463 ...
Страница 518: ...Index 498 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...