b. Tap Trigger Type and select a trigger type. This example uses Runt.
c. Tap Source and select a trigger source.
d. Tap Trigger When and select the condition on when to trigger on a runt signal.
e. Tap a Polarity icon to set the runt pulse polarity to detect (positive, negative, or either).
f.
Tap the Upper Threshold and Lower Threshold fields and set the levels that a signal must cross to be considered a
valid signal. Signals that don't cross both these thresholds are runt signals.
g. Tap anywhere outside the B Trigger Event menu to close it.
5. To trigger on a specific occurrence of the B trigger event:
a. Tap After the A Trigger Event is found: Trigger on the Nth Trigger Event button in the main Trigger menu.
b. Tap Where N is: and use the multipurpose knob to set the oscilloscope to trigger on the Nth occurrence of a B trigger
event.
6. To trigger on the B event after a specific time delay:
a. Tap After the A Trigger Event is found: Trigger on the 1st B Event button.
b. Tap After a Delay of: and use the multipurpose knob to set the desired delay time to wait before detecting and
triggering on a B trigger event. You can also double-tap the field and use the virtual keypad to enter a delay time.
Set up trigger on a parallel bus
Use this procedure to set up triggering on a parallel bus.
Use this procedure if you have already created a parallel bus.
1. Double-tap the Trigger badge.
2. Tap the Trigger Type field and select Bus from the list.
3. Tap the Source field and select the parallel bus on which to trigger.
4. Tap either the Binary or Hex Data boxes to enter the parallel bus data value, in either Binary or Hexadecimal format, on
which to trigger. The number of bits shown depends on the number of sources (channels) in the parallel bus.
a. Use multipurpose knob A to select the digit or digits to change.
b. Use multipurpose knob B to change the value of the selected digits.
Set up trigger on a serial bus
Use this procedure to set up triggering on a serial bus.
Use this procedure if you have already created a serial bus. Serial buses require purchasing and installing serial bus options.
See
Serial bus and trigger options
on page 14.
1. Double-tap the Trigger badge on the Settings bar.
2. Tap Trigger Type and select Bus from the list.
3. Tap Source and select a serial bus from the list.
4. Tap Trigger On and select what to trigger on from the list. The displayed fields and controls depend on the bus type and
Trigger On selection. Use these fields to trigger on a specific bus condition.
Advanced triggering
120
MSO54, MSO56, MSO58, MSO58LP, MSO64 Help
Содержание 6 series
Страница 24: ...Product documents and support 4 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 42: ...Options 22 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 54: ...Install your instrument 34 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 63: ...9 FlexChannel probe connectors Getting acquainted with your instrument MSO54 MSO56 MSO58 MSO58LP MSO64 Help 43 ...
Страница 84: ...Getting acquainted with your instrument 64 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 102: ...Configure the instrument 82 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 148: ...Advanced triggering 128 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 154: ...Zooming on waveforms 134 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 438: ...Waveform acquisition concepts 418 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 448: ...Waveform display concepts 428 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 483: ...Where σ is the Phase Angle Degree Measurement algorithms MSO54 MSO56 MSO58 MSO58LP MSO64 Help 463 ...
Страница 518: ...Index 498 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...