Turn On Time measurement: Configure panel.
Field or control
Description
Input Source
Selects the channel connected to the input side of the DUT.
Type
Selects the input to output power conversion type. Default is DC-DC.
Label
Sets the name of the measurement. You can use the default name, or double-tap in the field
and change the label using a connected keyboard or the virtual keyboard.
Maximum Voltage
Sets the maximum input voltage, in the range of 1 V to 500 V.
Input Trigger
Sets the input trigger level, in the range of 1 V to 500 V.
Frequency
Sets the DUT input AC signal frequency, in the range of 1 Hz to 1 MHz.
Available when Type = AC-DC
Maximum Time
Sets the maximum time window in which to capture the turn on time measurement (from the
initial state to a steady state), in the range of 1 μs to 500 s, with a resolution of 0.1 μs.
Number of Outputs (1, 2, 3, 4,
5, 6, or 7)
Sets the number of outputs to measure on the DUT (maximum of seven).
Output Source
Selects the channel source to use for the listed output.
Output Voltage
Sets the expected maximum voltage for each listed output, in the range of -6 kV to +6 kV, with a
resolution of 0.001 V. The measurement uses this value to set the vertical scale units for each
output waveform.
The instrument measures the actual maximum voltage on each output, and uses this value to
determine the output On/Off state levels, where On is ≥90% of the measured value, and Off is
≤10% of the measured value.
NOTE. The configuration value for maximum voltage and trigger level will not be remembered when it changes from AC-AC, AC-
DC, DC-AC, and DC-DC modes. It displays the default values.
Menus and dialog boxes
MSO54, MSO56, MSO58, MSO58LP, MSO64 Help
181
Содержание 6 series
Страница 24: ...Product documents and support 4 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 42: ...Options 22 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 54: ...Install your instrument 34 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 63: ...9 FlexChannel probe connectors Getting acquainted with your instrument MSO54 MSO56 MSO58 MSO58LP MSO64 Help 43 ...
Страница 84: ...Getting acquainted with your instrument 64 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 102: ...Configure the instrument 82 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 148: ...Advanced triggering 128 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 154: ...Zooming on waveforms 134 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 438: ...Waveform acquisition concepts 418 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 448: ...Waveform display concepts 428 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 483: ...Where σ is the Phase Angle Degree Measurement algorithms MSO54 MSO56 MSO58 MSO58LP MSO64 Help 463 ...
Страница 518: ...Index 498 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...