Other panel fields and controls.
Field or control
Description
Deskew
Sets or displays the probe deskew value.
Set to 0
Sets the probe deskew value to zero (0) seconds.
Multi-Channel
Opens a Deskew configuration menu that allows you to deskew multiple channels (two at a
time).
External Attenuation
Double-tap the numeric fields to set the external attenuation using the virtual keypad.
Set to Unity
Sets the external attenuation to unity.
Alternate Units
Toggles Alternate Units On or Off. Use to set custom vertical measurement units. The vertical
scale will show the entered measurement units.
Units
Double-tap to enter alternate units using the virtual keyboard. Only shown when Alternate
Units = On.
Ratio
Double-tap the numeric fields to set the ratio using the virtual keypad. Only shown when
Alternate Units = On.
Set to Unity
Sets the ratio to unity. Only shown when Alternate Units = On.
See also.
Probe Compensation configuration menu (analog channels Probe Setup panel)
on page 267
Deskew configuration menu (Other panel, Channel configuration menu)
on page 269
Deskew configuration menu (Other panel, Channel configuration menu)
Use the Deskew configuration menu to make display and measurement adjustments for analog probes that have differing
propagation delays. This is especially important when using a current probe in conjunction with a voltage probe.
To open the Deskew configuration menu:
1. Double-tap an analog Channel badge on the Settings bar to open the Channel configuration menu.
2. Tap the Other panel.
3. Tap the Multi-Channel button.
Use the controls in the Deskew menu to set the deskew parameters to recommended values, based on the nominal propagation
delay of supported probes. The oscilloscope automatically loads the nominal propagation delay values of TekVPI and TekProbe
II probes (TekProbe II probes require use of a TPA-BNC adaptor).
NOTE. This deskew menu does not actively test and adjust the probe delay between channels; it uses the delay values stored in
supported probes, or a custom propagation delay value that you enter, to set the propagation delay to zero between the
reference channel probe and one or more other probes.
To actively adjust probe delay using a signal, see
Deskew analog input channels - quick visual method
on page 78 and
analog input channels - measurement method
on page 79.
Menus and dialog boxes
MSO54, MSO56, MSO58, MSO58LP, MSO64 Help
269
Содержание 6 series
Страница 24: ...Product documents and support 4 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 42: ...Options 22 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 54: ...Install your instrument 34 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 63: ...9 FlexChannel probe connectors Getting acquainted with your instrument MSO54 MSO56 MSO58 MSO58LP MSO64 Help 43 ...
Страница 84: ...Getting acquainted with your instrument 64 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 102: ...Configure the instrument 82 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 148: ...Advanced triggering 128 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 154: ...Zooming on waveforms 134 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 438: ...Waveform acquisition concepts 418 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 448: ...Waveform display concepts 428 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 483: ...Where σ is the Phase Angle Degree Measurement algorithms MSO54 MSO56 MSO58 MSO58LP MSO64 Help 463 ...
Страница 518: ...Index 498 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...