Warning Message
Cause
Suggestion
Invalid Mask
Mask is not of the standard
format, cannot have closed
mask coordinates or
intersection of the coordinate
points
Recreate the mask from the SOA mask configure table such
that inner mask coordinates do not intersect.
No data in range
No data in the gated region.
Make sure that the input waveforms are valid.
Input > 80% DC
Input signal has higher DC
component.
Change the input signal to AC or add more AC component to
the DC signal.
Trigger Level is greater than
maximum voltage
Trigger level set is greater then
the maximum voltage.
Set the maximum voltage higher than the trigger level.
Output Voltage Level reached
before input trigger
Output is turned on before the
input or there is a spike in the
output voltage.
Make sure that the input and output voltage are not
interchanged, output is dependent on the input, and run the
measurement again.
Information messages displayed on the Power measurement badges
Information messages
Description
Actual Input Level used in calculation is n V.
In Turn On Time and Turn Off Time measurements, if the input
level is not equal to the maximum of input waveform (n), the
maximum of the input waveform is used for computation and
the same is displayed as the information message.
Actual Output x Level used in calculation is n V.
In Turn On Time and Turn Off Time measurements, if the
output x level is not equal to the maximum output x waveform
(x), the maximum of the output x waveform is used for
computation and the same is displayed as the information
message.
References
480
MSO54, MSO56, MSO58, MSO58LP, MSO64 Help
Содержание 6 series
Страница 24: ...Product documents and support 4 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 42: ...Options 22 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 54: ...Install your instrument 34 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 63: ...9 FlexChannel probe connectors Getting acquainted with your instrument MSO54 MSO56 MSO58 MSO58LP MSO64 Help 43 ...
Страница 84: ...Getting acquainted with your instrument 64 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 102: ...Configure the instrument 82 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 148: ...Advanced triggering 128 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 154: ...Zooming on waveforms 134 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 438: ...Waveform acquisition concepts 418 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 448: ...Waveform display concepts 428 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 483: ...Where σ is the Phase Angle Degree Measurement algorithms MSO54 MSO56 MSO58 MSO58LP MSO64 Help 463 ...
Страница 518: ...Index 498 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...