Other trigger types.
■
■
■
■
■
■
■
■
■
Rise/Fall Time Trigger configuration menu
Use the Rise/Fall Time trigger to trigger when the rise or fall time of a signal is less than, greater than, equal to, or not equal to a
specified time limit.
To open the Rise/Fall Time trigger configuration menu:
1. Double-tap the Trigger badge on the Settings bar.
2. Set the Trigger Type to Rise/Fall Time.
Settings panel (Rise/Fall Time Trigger configuration menu) fields and controls.
Field or control
Description
Source
Lists the source channel or waveform to use to trigger or search. Types that require multiple
inputs will replace this control with a different source definition control.
Trigger When
■
< Limit: A signal has a rise time less than the specified time limit.
■
> Limit: A signal has a rise time greater than the specified time limit.
■
= Limit: A signal has a rise time that is equal to the specified time limit (±5%).
■
≠ Limit: A signal has a rise time that does not equal (is greater than or less than) the
specified time limit (±5%).
Time Limit
Sets the time period condition to be met.
Slope
Sets the signal transition direction to detect. (rising, falling, or either direction).
Upper Threshold
Sets the upper threshold amplitude level through which the signal must pass to be considered a
valid transition.
Lower Threshold
Sets the lower threshold amplitude level through which the signal must pass to be considered a
valid transition.
Logic Qualification
When set to On enables logic qualification to further refine the trigger condition by setting the
required logic conditions on the source signals to generate a trigger event.
Define Inputs
Opens the Logic Qualification - Define Inputs configuration menu. Use this menu to set the
logic state, threshold levels, and logic operation of the input signals. See
Define Inputs configuration menu
on page 400.
Available when Logic Qualification = On.
Menus and dialog boxes
386
MSO54, MSO56, MSO58, MSO58LP, MSO64 Help
Содержание 6 series
Страница 24: ...Product documents and support 4 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 42: ...Options 22 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 54: ...Install your instrument 34 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 63: ...9 FlexChannel probe connectors Getting acquainted with your instrument MSO54 MSO56 MSO58 MSO58LP MSO64 Help 43 ...
Страница 84: ...Getting acquainted with your instrument 64 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 102: ...Configure the instrument 82 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 148: ...Advanced triggering 128 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 154: ...Zooming on waveforms 134 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 438: ...Waveform acquisition concepts 418 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 448: ...Waveform display concepts 428 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 483: ...Where σ is the Phase Angle Degree Measurement algorithms MSO54 MSO56 MSO58 MSO58LP MSO64 Help 463 ...
Страница 518: ...Index 498 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...