Switching Loss measurement: Configure panel.
Field or control
Description
Voltage Source
Selects the voltage source used to take the measurement. Tap the field to show the list of
available sources.
Current Source
Selects the current source used to take the measurement. Tap the field to show the list of
available sources.
Label
Sets the name of the measurement. You can use the default name, or double-tap in the field
and change the label using a connected keyboard or the virtual keyboard.
Type
Select SMPS / PFC/ Flyback:
■
SMPS: Select this option in case of signals without noise and ringing. The Vg source is not
required. Select Vg source (Source 3), in case of noisy signal.
■
PFC: Select this option when input DUT signals are from Power Factor Correction Circuit.
For this case, Vg source is mandatory.
■
Flyback: Select this option when input signals are ringing. This option does not require a
Vg source.
NOTE. Displayed REF levels shown in the Reference Levels panel after selecting Flyback
are default values and not the internally computed REF values used for measurement
calculations.
Gate Voltage (Vg)
Sets the Vg input source, which is a clean signal.
Available when Type = SMPS or PFC.
Vg Level Ton - Start
Selects the source of computation of the harmonic. In the voltage source, standard is always
None.
Available when Gate Voltage (Vg) ≠ None.
PWM Type
Select Fixed or Variable based on the varying pulse width of the switching signal.
Conduction calculation
Select MOSFET or BJT/IGBT semiconductor types.
R
DS
(on)
If MOSFET is selected, then R
DS
(on) is used to compute for conduction Loss.
Available when Conduction Calculation = MOSFET
V
CE
(sat)
If BJT/IGBT is selected, then V
CE
(sat) is used to compute for conduction loss.
Available when Conduction Calculation = BJT/IGBT.
Set On/Off Levels In:
Sets the REF levels for computation of the T
ON
and T
OFF
regions. Levels can be set in % or
absolute values.
T
off
-Stop Current Level
Sets the T
on
-Start and T
off
-Stop of the max switch current.
T
on
-Stop & T
off
- Start Voltage
Level
Sets the voltage level value for T
off
-Stop and T
on
. Can be entered as a percent or as a voltage,
depending on the setting of the Set On/Off Levels In: control.
Menus and dialog boxes
MSO54, MSO56, MSO58, MSO58LP, MSO64 Help
179
Содержание 6 series
Страница 24: ...Product documents and support 4 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 42: ...Options 22 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 54: ...Install your instrument 34 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 63: ...9 FlexChannel probe connectors Getting acquainted with your instrument MSO54 MSO56 MSO58 MSO58LP MSO64 Help 43 ...
Страница 84: ...Getting acquainted with your instrument 64 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 102: ...Configure the instrument 82 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 148: ...Advanced triggering 128 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 154: ...Zooming on waveforms 134 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 438: ...Waveform acquisition concepts 418 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 448: ...Waveform display concepts 428 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 483: ...Where σ is the Phase Angle Degree Measurement algorithms MSO54 MSO56 MSO58 MSO58LP MSO64 Help 463 ...
Страница 518: ...Index 498 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...