12713020-002-2 Revision D – April 2004
SSU-2000 User’s Guide 423
Stratum 1 Clock Module
Stratum 1 Clock Module
lists the software configuration options and factory defaults for the Type 1
Clock module
.
Clk Switch AR
On
On/Off
Input Switch
AutoReturn (AR)
AR (2)/AS On (1)/AS Off (0)
Input Selection Mode
Priority
Priority (0)/PQL (1)
Local Oscillator (LO)
On
On/Off
Frequency Offset
57 ppb
Constant
Elevation Time
86400 seconds
60 to 500000 seconds
Table E-3. Type 1 Clock Module Options/Defaults
Setting
Default
Range
Warmup Time
1200 seconds
900-3600 seconds
Min Tau
330 seconds
150 – 1200 seconds
Max Tau
450 seconds
150 – 1200 seconds
Clk Switch AR
On
On/Off
Input Switch
AutoReturn (AR)
AR (2)/AS On (1)/AS Off (0)
Input Selection Mode
Priority
Priority (0)/Pql (1)
Local Oscillator (LO)
OFF
On/Off
Alarm Elevation Time
86400 seconds
60 to 500,000 seconds;
0 disables alarm
Table E-2. Stratum 2E and 3E Clock Module Default Settings (Continued)
Setting
Factory Default
Range
Содержание SDU-2000
Страница 12: ...12 SSU 2000 User s Guide 12713020 002 2 Revision D April 2004 ...
Страница 18: ...18 SSU 2000 User s Guide 12713020 002 2 Revision D April 2004 ...
Страница 142: ...Clock Module Reference Data Communications Module 142 SSU 2000 User s Guide 12713020 002 2 Revision D April 2004 ...
Страница 172: ...Input Module Reference Data Input Adapter Panels 172 SSU 2000 User s Guide 12713020 002 2 Revision D April 2004 ...
Страница 222: ...Output Module Reference Data Output Adapter Panels 222 SSU 2000 User s Guide 12713020 002 2 Revision D April 2004 ...
Страница 390: ...SNMP Protocol 390 SSU 2000 User s Guide 12713020 002 2 Revision D April 2004 ...
Страница 420: ...I O Connector Grounding 420 SSU 2000 User s Guide 12713020 002 2 Revision D April 2004 ...
Страница 434: ...User Changes to Factory Defaults 434 SSU 2000 User s Guide 12713020 002 2 Revision D April 2004 ...