2. Description of the Piranha
Hardware and Specifications
2014-06/5.5C
17
Piranha & QABrowser Reference Manual
Note 1
: This is valid for a tube with 14° anode angle. The HVL for a 22° anode is typically 0,5 mm
lower (@ 80 kV, 3 mm TF).
Note 2
: All kerma and kerma rate ranges, inaccuracy, and resolution figures are valid for product
version 2 and higher of the Piranha.
Note 3
: The Kerma rate is calculated as the Kerma (Dose) divided by the Irradiation time. See
also
Waveforms and Triggers
.
Note 4
: The HVL range is valid if also the TF is within its specified range. For high TF at high kV
the HVL range may be limited by this.
Mammography
Parameter
RQ
Range
Inaccuracy
Resolution
kVp (standard)
Mo / 30 µm Mo
Mo / 25 µm Rh
Rh / 25 µm Rh
W / 50 µm Rh
3
W / 0.50 mm Al
5
Mo / 1.0 mm Al
W / 50 µm Ag
W / 75 µm Ag
W / 50 µm Rh (Gio)
W / 0.70 mm Al
W / 50 µm Ag (Sel)
4
W / 50 µm Rh (Sel)
4
M1
M3
M4
M6
3
M7
5
M8
M10
M11
M12
M15
M16
4
M17
4
18 – 49 kV
22 – 46 kV
25 – 49 kV
20 – 49 kV
20 – 48 kV
18 – 49 kV
20 – 40 kV
20 – 40 kV
22 – 35 kV
20 – 49 kV
22 – 39 kV
22 – 39 kV
±1.5 % or ±0.7 kV
±2 % or ±1 kV
±2 % or ±1 kV
±2 % or ±1 kV
±2 % or ±1 kV
±2 % or ±1 kV
±2 % or ±1 kV
±2 % or ±1 kV
±2 % or ±1 kV
±1.5 % or ±0.7 kV
±2 % or ±1 kV
±2 % or ±1 kV
4 digits
(10 V)
kVp (optional)
Mo / 30 µm Mo +
+ 2 mm Al
Mo / 2.0 mm Al
M1d
M2
25 – 35 kV
18 – 49 kV
±2 % or ±1 kV
±2 % or ±1 kV
4 digits
(10 V)
Irradiation time
0.1 ms – 2000 s
1 – 65535 pulses
±1 % or ±0.5 ms
±1 pulse
0.5 ms
1 pulse
Air kerma (Dose)
1
with wide range option
(WR)
5 µGy – 1500 Gy
25 nGy – 1500 Gy
3 µR – 150 kR
±5 %
±5 %
–
Air kerma rate
1
(Dose rate)
with wide range option
(WR)
10 µGy/s – 750 mGy/s
2
25 nGy/s – 750 mGy/s
2
30 µR/s – 86 R/s
1.8 mR/min – 5100 R/
min
±5 % or ±12 nGy/s
±5 % or ±1.5 µR/s
±5 % or ±0.1 mR/min
(for Irr. time >20 ms)
Typ. noise:
6 nGy/s
-Free run
-High Sensitivity
-Low Sensitivity
25 nGy/s – 20 mGy/s
0.25 µGy/s – 20 mGy/s
45 µGy/s – 750 mGy/s
±5 % or ±12 nGy/s
±5 % or ±12 nGy/s
±5 % or ±0.2
µGy/s
Typ. noise:
6 nGy/s
Note 1
: All kerma and kerma rate ranges, inaccuracy, and resolution figures are valid for product
version 2 and higher of the Piranha.
Note 2
: The Kerma rate is calculated as the Kerma (Dose) divided by the Irradiation time. See
also
Waveforms and Triggers
.
Note 3
: The M6 (
W / 50 µm Rh)
, M10 (
W / 50 µm Ag),
and M15 (
W / 0.70 mm Al
) calibrations are
suitable for the Hologic Selenia Dimensions and Fuji Amulet.
Note 4
: The M16 (
W / 50 µm Ag (Sel)
) and M17 (
W / 50 µm Rh (Sel)
) calibrations are suitable for
Hologic Selenia with W anode..
Note 5
: The M7 (
W / 0.5 mmAl
) calibration is suitable for Philips MicroDose Mammography
(Sectra).
82
82
Содержание Piranha
Страница 1: ...Reference Manual English Version 5 5C Piranha QABrowser RTI article number 9629050 10 ...
Страница 8: ...Introduction Chapter 1 ...
Страница 12: ...Description of the Piranha Chapter 2 ...
Страница 29: ...2 Description of the Piranha Hardware and Specifications 2014 06 5 5C 25 Piranha QABrowser Reference Manual ...
Страница 40: ...Description of the QABrowser Chapter 3 ...
Страница 77: ...Measurement Principles Theory Chapter 4 ...
Страница 91: ...Measurements with the Piranha System Chapter 5 ...
Страница 151: ...Optional Accessories Chapter 6 ...
Страница 154: ...6 Optional Accessories Piranha Panoramic Holder Piranha QABrowser Reference Manual 2014 06 5 5C 150 ...
Страница 155: ...Problems and Solutions Chapter 7 ...
Страница 161: ...Glossary Chapter 8 ...
Страница 181: ...Notes Piranha QABrowser Reference Manual 2014 06 5 5C Notes 177 ...