
5. Measurements with the Piranha System
Dental and Panoramic Dental
2014-06/5.5C
131
Piranha & QABrowser Reference Manual
the kV waveform is displayed. To see more of the waveform, change the
kV range.
5.6.3
Panoramic Systems
Use the same procedure as for the normal dental measurement but select
Panoramic
Dental
instead of radiographic and select
Tube Voltage
as parameter. A 200 ms delay
is default.
1. Set up the Piranha and the handheld computer according to the description in
Setting
Up the Piranha
. Selecting one parameter mode enables you to see the measured
values from a distance of several meters.
2. Mount the Piranha and power on the
Piranha.
3. The real-time display for Tube voltage
is now displayed.
4. It is strongly advised to make Position
Check to confirm that the detector
area is uniformed irradiated if you
want reliable kVp readings. Select
the
Check[C]
filter and leave the
room to make the first panoramic
scan.
5. Start the exposure. You do not need to use the whole scan exposure to make a
12
Содержание Piranha
Страница 1: ...Reference Manual English Version 5 5C Piranha QABrowser RTI article number 9629050 10 ...
Страница 8: ...Introduction Chapter 1 ...
Страница 12: ...Description of the Piranha Chapter 2 ...
Страница 29: ...2 Description of the Piranha Hardware and Specifications 2014 06 5 5C 25 Piranha QABrowser Reference Manual ...
Страница 40: ...Description of the QABrowser Chapter 3 ...
Страница 77: ...Measurement Principles Theory Chapter 4 ...
Страница 91: ...Measurements with the Piranha System Chapter 5 ...
Страница 151: ...Optional Accessories Chapter 6 ...
Страница 154: ...6 Optional Accessories Piranha Panoramic Holder Piranha QABrowser Reference Manual 2014 06 5 5C 150 ...
Страница 155: ...Problems and Solutions Chapter 7 ...
Страница 161: ...Glossary Chapter 8 ...
Страница 181: ...Notes Piranha QABrowser Reference Manual 2014 06 5 5C Notes 177 ...