5. Measurements with the Piranha System
Fluoroscopy and Pulsed Fluoroscopy
2014-06/5.5C
105
Piranha & QABrowser Reference Manual
4. Select
I.I. input dose rate
from the menu.
5. The Select Detector screen is now displayed. Select External.
6. Tap
Select
.
7. Place the Piranha Dose Probe in front of the image intensifier but outside the
measuring field for the mA feedback loop. You may use the optional detector rod that
can be attached to Piranha Dose Probe to position the detector without risk for
hazardous X-ray exposure. Observe the image on the monitor.
8. The real-time display is now displayed. Set the generator. Tap
Reset
.
Since the external dose probe (Piranha Dose Probe) is not sensitive to
back scatter, a lower value compared to a transmission ion chamber is
typically detected (typically in the range of 5 - 20 %).
You may use the
beam correction factor
to make automatic corrections. The
beam correction factor
may also be stored permanently in a
Favourite
for easy
access.
Содержание Piranha
Страница 1: ...Reference Manual English Version 5 5C Piranha QABrowser RTI article number 9629050 10 ...
Страница 8: ...Introduction Chapter 1 ...
Страница 12: ...Description of the Piranha Chapter 2 ...
Страница 29: ...2 Description of the Piranha Hardware and Specifications 2014 06 5 5C 25 Piranha QABrowser Reference Manual ...
Страница 40: ...Description of the QABrowser Chapter 3 ...
Страница 77: ...Measurement Principles Theory Chapter 4 ...
Страница 91: ...Measurements with the Piranha System Chapter 5 ...
Страница 151: ...Optional Accessories Chapter 6 ...
Страница 154: ...6 Optional Accessories Piranha Panoramic Holder Piranha QABrowser Reference Manual 2014 06 5 5C 150 ...
Страница 155: ...Problems and Solutions Chapter 7 ...
Страница 161: ...Glossary Chapter 8 ...
Страница 181: ...Notes Piranha QABrowser Reference Manual 2014 06 5 5C Notes 177 ...