5. Measurements with the Piranha System
Radiography
Piranha & QABrowser Reference Manual
2014-06/5.5C
96
When the dose value has been reduced to
less than half the HVL and total filtration
value are calculated.
Tap
Graph
to view a graphical presentation
of the result.
It is recommended to use the built-in HVL application (or Ocean) to evaluate HVL.
5.2.3.1
Introduction 2
The Piranha system can, depending on model, measure up to eight parameters
simultaneously plus three waveforms from a single exposure:
kVp
Dose and dose rate
Exposure time
mAs and mA
Estimated total filtration and determined waveform type
kV waveform
Dose rate waveform
mA waveform