5. Measurements with the Piranha System
Dental and Panoramic Dental
Piranha & QABrowser Reference Manual
2014-06/5.5C
128
You can either select the tube voltage as single parameter or together with dose, dose
rate, and exposure time. As complementary information an estimation of the total
filtration in the beam and type of waveform are made. This features use the kVp filter
R1[4] that also is the default kV range 55 - 105 kV when the instrument is turned on.
The displayed dose value has very little energy dependence since it is automatically
compensated for each exposure since the kV, estimated filtration and the waveform are
measured.
5.6.1
kVp, Time, Dose, and Dose Rate
Use the same procedure as for the normal radiography measurement but select
Dental
instead of radiography and select
Tube voltage
as parameter. A 200 ms delay is
default.
1. Set up the Piranha and the handheld computer according to the description in
Setting
Up the Piranha
. Selecting one parameter mode enables you to see the measured
values from a distance of several meters.
If you want to measure the exposure time in
number of pulses you can do so.
12
Содержание Piranha
Страница 1: ...Reference Manual English Version 5 5C Piranha QABrowser RTI article number 9629050 10 ...
Страница 8: ...Introduction Chapter 1 ...
Страница 12: ...Description of the Piranha Chapter 2 ...
Страница 29: ...2 Description of the Piranha Hardware and Specifications 2014 06 5 5C 25 Piranha QABrowser Reference Manual ...
Страница 40: ...Description of the QABrowser Chapter 3 ...
Страница 77: ...Measurement Principles Theory Chapter 4 ...
Страница 91: ...Measurements with the Piranha System Chapter 5 ...
Страница 151: ...Optional Accessories Chapter 6 ...
Страница 154: ...6 Optional Accessories Piranha Panoramic Holder Piranha QABrowser Reference Manual 2014 06 5 5C 150 ...
Страница 155: ...Problems and Solutions Chapter 7 ...
Страница 161: ...Glossary Chapter 8 ...
Страница 181: ...Notes Piranha QABrowser Reference Manual 2014 06 5 5C Notes 177 ...