5. Measurements with the Piranha System
CT
2014-06/5.5C
135
Piranha & QABrowser Reference Manual
3. Start the QABrowser and select
CT
from the
Select type of measurement menu.
4. Select
Tube voltage
.
5. The real-time display for tube voltage is now
shown. At the bottom of the screen you can
select kV range and calibration. Three different
calibrations are available:
C1 = W/3.0 mm Al
C2 = W/3.0 mm Al + 0.25 mm Cu (optional)
C3 = W/3.0 mm Al + 1.2 mm Ti (optional)
In earlier software versions, the Piranha
could not measure total filtration in the CT
application. Therefore, an extra radiation quality
(C2) corresponding to measurements inside a
phantom, was necessary in order to get a
correct kV. Now this is taken care of
automatically, since the total filtration is
measured also for CT.
2: C3 is a special calibration for kV only. You
cannot measure TF and HVL with it. For work-
around see FAQ at RTI's webpage.
Содержание Piranha
Страница 1: ...Reference Manual English Version 5 5C Piranha QABrowser RTI article number 9629050 10 ...
Страница 8: ...Introduction Chapter 1 ...
Страница 12: ...Description of the Piranha Chapter 2 ...
Страница 29: ...2 Description of the Piranha Hardware and Specifications 2014 06 5 5C 25 Piranha QABrowser Reference Manual ...
Страница 40: ...Description of the QABrowser Chapter 3 ...
Страница 77: ...Measurement Principles Theory Chapter 4 ...
Страница 91: ...Measurements with the Piranha System Chapter 5 ...
Страница 151: ...Optional Accessories Chapter 6 ...
Страница 154: ...6 Optional Accessories Piranha Panoramic Holder Piranha QABrowser Reference Manual 2014 06 5 5C 150 ...
Страница 155: ...Problems and Solutions Chapter 7 ...
Страница 161: ...Glossary Chapter 8 ...
Страница 181: ...Notes Piranha QABrowser Reference Manual 2014 06 5 5C Notes 177 ...