5. Measurements with the Piranha System
Mammography
2014-06/5.5C
119
Piranha & QABrowser Reference Manual
1. From the dose real-time display tap the
Appl
button.
2. Select
HVL
.
3. The HVL application appears.
4. Perform exposures and add filter according
to the information in the HVL application.
5. When the dose value has been reduced to
less than half the HVL is calculated and the
graph can be selected.
5.5.6
Quick-HVL
Quick-HVL
The Piranha is able to measure HVL for mammography in one shot. Quick-HVL is
available for measurements with or without compression paddle. If the parameter "
All
w. HVL
" is chosen, the QABrowser will display quick-HVL values for every
measurement.
5.5.7
Mammo Compensations and Corrections
Here various corrections and compensations are described, that are of special
importance for mammography.
Содержание Piranha
Страница 1: ...Reference Manual English Version 5 5C Piranha QABrowser RTI article number 9629050 10 ...
Страница 8: ...Introduction Chapter 1 ...
Страница 12: ...Description of the Piranha Chapter 2 ...
Страница 29: ...2 Description of the Piranha Hardware and Specifications 2014 06 5 5C 25 Piranha QABrowser Reference Manual ...
Страница 40: ...Description of the QABrowser Chapter 3 ...
Страница 77: ...Measurement Principles Theory Chapter 4 ...
Страница 91: ...Measurements with the Piranha System Chapter 5 ...
Страница 151: ...Optional Accessories Chapter 6 ...
Страница 154: ...6 Optional Accessories Piranha Panoramic Holder Piranha QABrowser Reference Manual 2014 06 5 5C 150 ...
Страница 155: ...Problems and Solutions Chapter 7 ...
Страница 161: ...Glossary Chapter 8 ...
Страница 181: ...Notes Piranha QABrowser Reference Manual 2014 06 5 5C Notes 177 ...