![RADWAG MAS 1. Y Скачать руководство пользователя страница 94](http://html1.mh-extra.com/html/radwag/mas-1-y/mas-1-y_user-manual_781002094.webp)
- 94 -
temperature and humidity values is displayed. For <Ambient condition> parameter set
to <ONLINE> parameter a window for confirmation of values taken from ambient
conditions module is displayed.
CAUTION:
If <Ambient conditions> parameter for the differential weighing mode is set to <No>
value, the software skips this part of the process.
Press <
> button to confirm ambient conditions values, differential weighing home
screen is displayed, you are prompted to unload the weighing pan.
Unload the weighing pan and press <
> button. The balance searches samples list
again in order to find the first sample for which weighing <A> is still to be carried out.
Follow the above procedure and carry out weighing of the remaining samples.
In order to abort the process press <
> button.
Measurement of series samples can be carried under the following procedures: weighing A,
weighing T or weighing T+A. Each of the series samples features status informing you on a
process stage for a given sample. Upon entering the series database and selecting a
particular series for preview you are provided with list of samples recorded for the series.
Status overview:
•
weighing A, measurements completed for <Weighing A> or <Weighing T+A>
procedure,
•
weighing T, measurements completed for <Weighing T> procedure.
In order to preview detailed information on a sample, press field with its name.
Sample window contains the following data: