- 72 -
6. Load the pan with a pycnometer, wait for a stable measurement result and press
<
> button.
7. Take the pycnometer off the pan, fill it with tested substance, follow the pycnometer
guidelines.
8. Load the pan with a pycnometer, wait for a stable measurement result and press
<
> button.
9. Density result is displayed.
10. Press <
> button to complete the procedure.
CAUTION:
Press <
> button to start new density determination process with the same settings.
20.7. Mode Related Settings
The supplementary settings enable you to adjust the working mode to your needs and
requirements.
o
Request sample number:
NO
- sample number not required; single measurement.
YES
- sample number required; series of measurements.
o
Measuring unit: [g/cm
3
], [kg/m
3
], [g/l]; selected unit is valid for all density
determining options and for final indication printouts.
o
Assign density to a product: YES/NO; the software automatically assigns the
value of determined density of solid or liquid to <Density> parameter in a product
record of the database of products. To use this function first enter products
database and select a product for which the density procedure is carried out. Next
start density determining process (if density has already been determined for this
product, it is replaced with new value).
o
Tare mode
o
Printout/Enter mode
o
Printout
To find out more about the remaining functions refer to section
15.7
of this user manual,
Additional Parameters for a Weighing Process
.
20.8. Printouts
Printouts option enables you to determine particular standard printout content and to define a
non-standard printout.
Standard printout
Standard printout comprises four internal sections containing different variables.
Variable settings: YES - print, NO – not print.
Settings for header, weighing and footer printouts are to be found in section 14.5. For
density printout template read below.
You can design content of a report for each of carried out procedures. Upon pressing
<Density printout template> parameter, a window is opened where you can set content of a
particular report.