- 75 -
loading the pane with the next object at the moment when the indication exceeds the set
threshold value.
•
Threshold
Value expressed in mass units. In order to trigger the measurement, the indicated value has
to be greater than the threshold value.
3. Set parameters and return to weighing.
4. Load the weighing pan with an object to be weighed, press <
> button.
5. “Freezed” measurement result is displayed.
6. In order to proceed to the next measurement press <
> button:
- in case of manual operation mode, press <
> button.
- in case of automatic operation mode, unload the weighing pan and load it with the
new object to be weighed.
21.2. Mode Related Settings
To find out more about the remaining functions refer to section
15.7
of this user manual
Additional Parameters for a Weighing Process
.
22. FORMULATIONS
<
Formulations>
is a working mode allowing you to prepare multi-ingredient mixture.
The whole process is carried out automatically.
While making mixtures you can:
use formulations database
which stores formulations; the program helps you
dispense particular amount of formulation ingredients, this is done by means of
respective on-screen messages.
make mixture on your own, without using formulations database, with this option you
control dispensed ingredients, their sequence and quantity by yourself.
If you want to use formulations database, first you have to create recipe (make formulation),
next apply it. In order to create formulation, enter <Formulations> database. For detailed
procedure informing you how to create formulation, read later sections of this user manual.
Formulations mode screen:
1. Setup – press to access balance menu.
2. Print header.
3. Print footer.
4. Formulation – press to select formulation from
the database.
5. Formulation multiplier (formulation activated
when <NO> value is set for <Multiplier’s
editing> parameter).
6. Target mass
7. Formulation using out of database ingredients
22.1. Mode Related Settings
The supplementary settings enable you to adjust the working mode to your needs and
requirements. To access the settings follow the below procedure.
Содержание MAS 1. Y
Страница 1: ...WEIGHING MODULES MAS Y MPS Y MAS 1 Y USER MANUAL IMMU 25 01 08 18 EN www radwag com...
Страница 9: ...9 1 GENERAL INFORMATION 1 1 Dimensions MAS Module MPS Module...
Страница 10: ...10 MAS 1 Module...
Страница 200: ...200...
Страница 201: ...201...