- 175 -
Command overview:
Command gives information about currently set filter.
Format:
FIG<CR><LF>
Response options:
FIG_x_OK<CR><LF>
- command carried out, response: currently set filter
FIG_I <CR><LF>
- command understood but not accessible at this
moment
x
- parameter, symbol of currently set filter
Example:
Command:
FIG<CR><LF>
- give current filter
Response:
FIG_2_OK<CR><LF> - currently set filter: average
Set value release
Format:
ARS_n <CR><LF>
Response options:
ARS_OK <CR><LF>
- command carried out
ARS_E <CR><LF>
- error in-course of command execution, no parameter or incorrect
format
ARS_I <CR><LF>
- command understood but not accessible at this moment
n
- parameter, decimal value determining value release options
n
→
1 – fast
2 – fast + reliable
3 – reliable
CAUTION:
The numbering is assigned to a particular value release option and it is identical for all
balance types.
If for a particular balance type, last digit settings are assigned to a working mode, the
command changes the current mode's settings.
Example:
Command:
ARS_2<CR><LF>
- set value release parameter to fast+reliable
option
Response:
ARS_OK<CR><LF>
- fast+reliable option set
Give current value release setting
Command overview:
Command gives information about current value release setting.
Format:
ARG<CR><LF>
Response options:
ARG_x_OK<CR><LF>
- command carried out, response: current value release
setting
ARG_I <CR><LF>
- command understood but not accessible at this
moment
x
- parameter, symbol of current value release setting
Example:
Command: ARG<CR><LF>
- give current value release setting
Response: ARG_2_OK<CR><LF>
- current value release setting: fast+reliable
Set last digit