- 158 -
Example:
You can quickly modify ‘filter’ and ‘value release’ settings. To do that, select results for
particular settings that are to be saved. Press <Activate>.
Report example:
-------- Autotest Filter: Report --------
Balance type XA 4Y
Balance S/N 442566
Operator Hubert
Software ver. NL1.6.5 S
Date 2015.05.07
Time 09:34:48
----------------------------------------
Reading unit 0.0001/0.0001 g
Internal weight mass 148.9390 g
Temperature: Start 25.26 °C
Temperature: Stop 25.66 °C
----------------------------------------
Filter Very fast
Value release Fast
Repeatability 0.0042 g
Stabilization time 4.505 s
.
.
.
Filter Very slow
Value release Reliable
Repeatability 0.0207 g
Stabilization time 5.015 s
Signature
........................................