Procedure:
Upon ‘Autotest Filter’ function activation, the process starts automatically. Progress bar is
displayed. Upon process completion, a summary is displayed, current filter settings are
provided. The report can be printed.
You can abort the process at any moment. To do that press <X> button.
AUTOTEST GLP
The test controls repeatability of loading the internal weight and determining error of
indication with reference to module’s maximum capacity.
Procedure:
•
Two times loading of the internal weight, followed by 10 times loading of the
internal weight.
•
Adjust the weighing instrument,
•
Calculating and recording the value of standard deviation.
•
In weighing instrument with automatically opened doors, door opening and
closing test is carried out.
Autotest GLP function enables:
Report display and printout,
Record of a report containing basic balance data, ambient conditions values and
test results.
Test results:
*Deviation for maximum capacity.
*Value of repeatability expressed as standard deviation.
*Door test result (positive / negative).
Report example:
.....................................................
--------- Autotest GLP: Report --------
Balance type XA 4Y
Balance S/N 400010
Operator Admin
Software ver. L0.0.21 S
Date 2012.01.16
Time 09:17:16
----------------------------------------
Measurements quantity 10
Reading unit 0.0001 g
Internal weight mass 140.094 g
Filter Average
Value release Fast & reliable
----------------------------------------
Deviation for Max. -0.0118 g
Repeatability 0.00088 g
Signature
........................................