- 24 -
Press to search for a particular record in a database by name
Press to search for a particular record in a database by code
Press to print out a particular record from a database
Press to clear the editing field
Press to enable/disable an on-screen keyboard
Press to read printout template from a file format *.lb (function
key is active on inserting a USB flash drive to terminal’s USB
port)
Press to select variables for a printout from a list
Press to move (exit) one level up
9.2. Return to the Weighing Mode
Modifications introduced into balance memory, are automatically saved upon return to
weighing operation.
Procedure:
•
Press
button repeatedly, keep pressing until you see the home screen.
•
Press
button, the home screen is displayed immediately.
10. ADJUSTMENT
The module features automatic internal adjustment system which ensures correct
measurement accuracy. <ADJUSTMENT> menu contains functions controlling operation
of module adjustment process.
10.1. Internal Adjustment
Internal adjustment is carried out by means of an internal adjustment weight. <Internal
adjustment> button, when pressed, automatically triggers adjustment process. Upon
adjustment process completion, a respective message is displayed informing on the
process status.
CAUTION:
Balance adjustment procedure requires stable conditions (no air drafts, no vibrations), in
the course of adjustment the weighing pan must be unloaded
.
10.2. External Adjustment
External adjustment is carried out using an external mass standard of the right accuracy
and weight value, which value depends on weighing device model and capacity. The
external adjustment procedure is a semi-automatic process, particular stages of which are
signalled with on-screen prompts.