- 85 -
Functions:
−
Measurements quantity
: designed to enable determining measurements quantity
for tested pipette volume (valid for each tested volume in case of a pipette with
adjustable volume),
−
Request serial number
: YES/NO. YES – a window with an on-screen keyboard is
displayed, use it to enter serial number of a pipette.
−
Operation with adherence to ISO 8655
: YES/NO. YES – the software
automatically takes error values accordant to ISO 8655 standard (if other error
values have been defined for a particular pipette, then on selecting <Operation with
adherence to ISO 8655> parameter the specified error values are neglected and
replaced by those from ISO standard).
−
Upload ambient conditions parameters from THB module
: YES/NO. YES – the
software automatically acquires and saves ambient conditions’ parameters from a
connected THB sensor at a right time. NO – you have to read ambient conditions
(i.e. temperature, humidity and pressure) from external sensors and enter them
manually both before and after calibration.
−
Result control, Tare mode, Automatic footer printout, Printout mode, Printout
:
for information on how to operate these functions read section
15.7
.
24.2. Pipettes Calibration – Quick Access Buttons
Each working mode features automatically displayed set of default on-screen quick access
buttons. The set can be modified, i.e. you can add out-of-list quick access buttons to it. Such
operation requires particular permissions level.
24.3. Adding a Pipette to Pipettes Database
Database of pipettes contains list of pipettes’ names and other data including tested volumes
and error values for a specific volume. When adding a pipette to the database, first specify
its name, and then add remaining.
The software is intuitive and you are guided by means of respective messages. Pipettes
database menu has been designed to enable adding formulations.
Procedure:
•
Enter <
Databases> submenu and press <
Pipettes > button.
•
Press
Add> button to add a new pipette.
A new record is automatically added to the database, it is opened in an editing mode. Enter
respective data.
List of parameters defined for a pipette:
1. Name: press to enter pipette name.
2. Code: press to enter pipette code.
3. Model: press to enter model name.
4. Tip: press to enter tip name
5. Volume type: ADJUSTABLE/FIXED.
6. Nominal volume: pipette’s nominal volume value.
7. Minimum volume: pipette’s minimum volume value, for fixed volume pipette enter
<0>;
8. Channels quantity: specify number of pipette’s channels, for a single channel
pipette enter <1>;