4. Subsequent control - a field for entering expiry date of the minimum sample weight. 2
weeks before the expiry date, a pictogram of a clock is displayed next to the status
pictogram. Contact RADWAG representative to carry out necessary changes in
minimum sample weight settings.
5. Thresholds – option that enables entering data on minimum sample weight mass and
range of packaging mass (tare) with applicable criteria:
Tare
– maximum tare value, for which the minimum sample weight is obligatory.
You can enter three characteristic quantities: 0.000g, optional mass within weighing
instrument's measuring range, and maximum capacity (see: examples below).
Minimum mass –
value of the minimum sample weight determined for a particular
weighing device on site with use of respective method.
Example no. 1 XA 220.4Y balance, d=0.0001 g:
List of MSW values determined for the following packaging values:
No.
Tare value
Minimum
sample
weight
Overview
1
0.0000 g
0.1000 g
Minimum sample weight refers to all net weight
values when weighing is carried out without use
of a tare container (<TARE> button NOT used).
2
10.0000 g
1.0000 g
Minimum sample weight refers to all net weight
values when weighing is carried out using a tare
container of mass ranging from 0.0001 g to
10.0000 g inclusive (<TARE> button used).
3
50.0000 g
2.5000 g
Minimum sample weight refers to all net weight
values when weighing is carried out using a tare
container of mass ranging from 10.0001 g to
50.0000 g inclusive (<TARE> button used).
4
200.0000 g
4.0000 g
Minimum sample weight refers to all net weight
values when weighing is carried out using a tare
container of mass ranging from 50.0001 g to
200.0000 g inclusive (<TARE> button used).
Example no. 2 XA 220.4Y balance:
No.
Tare value
Minimum
sample
weight
Overview
1
220.0000 g
0.5000 g
Minimum sample weight refers to all net weight
values when weighing is carried out using a tare
container of mass value comprised within the
whole weighing range (<TARE> button used),
and to all net weights that are weighed without a
packaging (the <TARE> button not used).
Example no. 3 XA 220.4Y balance:
No.
Tare value
Minimum
sample
weight
Overview
1
0.0000 g
0.2500 g
Minimum sample weight refers to all net weight
values when weighing is carried out without use
of a tare container (<TARE> button NOT used).
In case of such settings, the minimum sample