![RADWAG MAS 1. Y Скачать руководство пользователя страница 59](http://html1.mh-extra.com/html/radwag/mas-1-y/mas-1-y_user-manual_781002059.webp)
- 59 -
16.6. Single Part Mass Update in a Database
Determined single part mass can be assigned to a product in the database of products.
This option is applied while using ACAI (Automatic Accuracy Correction) function in order
to determine part mass with high accuracy.
Procedure:
1. Determine single part mass.
2. Press <
Products> database.
3. Press and hold name of a product for which mass value is to be updated.
4. Context menu opens.
5. Press
<Assign reference sample mass>
button, reference sample mass for the
product is recorded as
<Mass>
.
16.7. Parts Counting Procedure
First, it is necessary to obtain weight value of a single piece. Select one of the available
options:
−
Enter weight value for a single part (section 16.3), next put parts to be counted on the
weighing pan, parts quantity is displayed.
−
Determine mass of a single part from a given quantity of parts (section 16.4.). ACAI
function
pictogram is displayed (providing that the function is active). Next put
parts to be counted on the weighing pan, parts quantity is displayed.
−
Acquire weight value for a single part from the database (section 16.5). To do it, you
have to select respective product. Now put parts to be counted on the weighing pan,
parts quantity is displayed.
Caution: It is necessary to tare any additional elements (packaging) prior to counting
process start.
16.8. Checkweighing for Parts Counting Mode
Parts counting process can be aided by checkweighing function, i.e. control whether
indication is within the set thresholds.
Checkweighing
requires setting values for two thresholds:
−
Min threshold [Min= … pcs.]
−
Max threshold [Max= … pcs.]
and setting YES value for Bar graph. The Bar graph displays relation: CURRENT PARTS
QUANTITY/CHECKWEIGHING THRESHOLDS.
In order to define Min and Max thresholds’ values open Products database and edit the
Product record or use <
Checkweighing thresholds> quick access button.
Procedure:
•
Press grey workspace, next select <Information> parameter.
•
Set Bar graph to YES value, return to parts counting mode.
Содержание MAS 1. Y