- 148 -
32.3.1. Barcode Scanner Port
Procedure:
•
Enter
<
Peripherals>
parameter and go to
„
Barcode scanner
/
Port”
parameter. Select respective option.
Ports for balance – scanner communication:
−
RS 232 (COM1),
−
RS 232 (COM2),
−
USB
32.3.2. Prefix/Suffix
You can edit
<
Prefix>
and
<
Suffix>
to synchronize mass comparator’s software
with barcode scanner.
CAUTION:
In RADWAG-adopted standard, the prefix is 01 sign (byte) hexadecimal format, the suffix is
0D sign (byte) hexadecimal format.
For detailed description of 'module' - 'barcode scanner' communication read APPENDIX E of
this user manual.
Procedure:
•
Enter
<
Barcode scanner>
submenu,
•
Select
<
Prefix>
parameter and enter hexadecimal value using the on-screen
keyboard. Press
button to confirm.
•
Select
<
Prefix>
parameter and enter hexadecimal value using the on-screen
keyboard. Press
button to confirm.
32.3.3. Field Selection
Parameter allowing you to specify, for which field the search is to be carried out in particular
databases.
Procedure:
•
Enter <
Peripherals>
submenu,
•
Select
„
Barcode scanner
/
Field selection”
. List of fields is displayed:
Product
Series
Operator
Universal variable 1
Customer
Universal variable 2