- 49 -
Each
– printout of all measurements (stable and unstable), for
verified balance only stable measurements are printed out (as
in <Each stable> setting).
•
AUTOMATIC MODE
Never – printout disabled.
First stable
- first stable measurement is recorded upon loading of the
weighing pan with a load; record of the next stable
measurement result takes place only if the following conditions
are met: weighing pan gets unloaded, measurement value
lower than value of set threshold is displayed, a new load is
placed on the weighing pan.
Last stable - last stable measurement, registered prior unloading, is
recorded. The record takes place only if the following conditions are met:
weighing pan gets unloaded, measurement value lower than value of set
threshold is displayed
With interval
– select this value to start automatic printout and record of
indications in WEIGHINGS database, carried out in a cyclic
manner in a specified time intervals. The interval is set in [min]
in <INTERVAL> parameter. Interval range is 1 - 9999 min.
CAUTION:
Each weighing result is printed and recorded (stable and unstable for a non-verified
balance, stable for a verified balance).
Automatic operation with interval starts at the moment of switching the function on.
The first stable weighing result of value greater than THRESHOLD value is printed
and recorded as the first measurement. The following measurements are printed
with frequency set in INTERVAL parameter. To stop the automatic operation with
interval, switch the option off.
THRESHOLD:
Mass value obligatory for operation of automatic printout, set in grams.
INTERVAL
Record frequency for automatic operation with interval
−
Printout
Printout type related to a working mode. Printout is activated by pressing PRINT
key on balance panel.
Available options:
•
STANDARD PRINTOUT:
Enables declaring content of the following printouts: HEADER, WEIGHINGS
and FOOTER.
Select <YES> parameter for the components that are to be printed on
pressing printout activating function key.
•
NON-STANDARD PRINTOUT:
Enables selecting one of database-stored non-standard printouts form a list
of <PRINTOUTS> menu, or designing a unique printout which is
automatically added to the database.
Caution: For detailed instruction refer to section 14.5 of this user manual.
−
Air buoyancy compensation
Parameters allowing you to switch on the compensation and to enter weighed
sample density data and air density data.
CAUTION:
Function valid for weighing mode exclusively.
For detailed information regarding the operation and settings refer to the previous
section of this user manual.