- 166 -
Response format:
1 2
3
4
5
6
7-15
16
17 18 19 20 21
S I
space
stability
marker
space character mass space
unit
CR LF
Example:
S I CR LF
- command sent from a computer
S I _ ? _ _ _ _ _ _ _ 1 8 . 5 _ k g _ CR LF
- command carried out, immediate response:
mass value in basic measuring unit
where:
_
- space
Send stable measurement result in current measuring unit
Format:
SU CR LF
Response options:
SU_A CR LF
- command understood and in progress
SU_E CR LF
- time limit exceeded while waiting for stable measurement result
SU_I CR LF
- command understood but not accessible at this moment
MASS FRAME
- command carried out, immediate response: mass value in basic measuring
unit
Response format:
1 2
3
4
5
6
7-15
16
17 18 19
20
21
S U space
stability
marker
space character mass space
unit
CR LF
Example:
S U CR LF
- command sent from a computer
S U _ A CR LF
- command understood and in progress
S U _ _ _ - _ _ 1 7 2 . 1 3 5 _ N _ _ CR LF
- command carried out, immediate response:
mass value in current measuring unit.
where:
_
- space
Immediately send measurement result in current measuring unit
Format:
SUI CR LF
Response options:
SUI_I CR LF
- command understood but not accessible at this moment
MASS FRAME - immediate response: mass value in basic measuring unit
Response format:
1 2 3
4
5
6
7-15
16
17 18 19 20 21
S U I
stability
marker
space
character mass
space
unit
CR LF
Example:
S U I CR LF
- command sent from a computer
S U I ? _ - _ _ _ 5 8 . 2 3 7 _ k g _ CR LF
- command carried out, immediate response:
mass value in basic measuring unit
where:
_
- space
Switch on continuous transmission in basic measuring unit
Format:
C1 CR LF
Response options:
C1_I CR LF
- command understood but not accessible at this moment
Содержание MAS 1. Y
Страница 1: ...WEIGHING MODULES MAS Y MPS Y MAS 1 Y USER MANUAL IMMU 25 01 08 18 EN www radwag com...
Страница 9: ...9 1 GENERAL INFORMATION 1 1 Dimensions MAS Module MPS Module...
Страница 10: ...10 MAS 1 Module...
Страница 200: ...200...
Страница 201: ...201...