T_I CR LF
- command understood but not accessible at this moment
Give tare value
Format:
OT CR LF
Response:
OT_TARE CR LF
- command carried out
Response format:
1
2
3
4-12
13
14
15
16
17
18
19
O
T
space
tare
space
unit
space
CR
LF
Tare
- 9 characters, right justification
Unit
- 3 characters, left justification
CAUTION:
Tare value is always given in adjustment unit.
Set tare
Format:
UT_TARE CR LF
, where
TARE
- tare value.
Response options:
UT_OK CR LF - command carried out
UT_I CR LF
- command understood but not accessible at this moment
ES CR LF
- command not recognised (tare format incorrect)
CAUTION:
Use dot in tare format as decimal point.
Send stable measurement result in basic measuring unit
Format:
S CR LF
Response options:
S_A CR LF
- command understood and in progress
S_E CR LF
- time limit exceeded while waiting for stable measurement result
S_I CR LF
- command understood but not accessible at this moment
MASS FRAME
- command carried out, immediate response: mass value in basic measuring
unit
Response format:
1
2-3
4
5
6
7-15
16
17 18 19 20 21
S space
stability
marker
space character mass
space
unit
CR LF
Example:
S CR LF
- command sent from a computer
S _ A CR LF
- command understood and in progress
S _ _ _ _ - _ _ _ _ _ _ 8 . 5 _ g _ _ CR LF
- command carried out, immediate response:
mass value in basic measuring unit
where:
_
- space
Immediately send measurement result in basic measuring unit
Format:
SI CR LF
Response options:
SI_I CR LF
- command understood but not accessible at this moment
MASS FRAME - immediate response: mass value in basic measuring unit