- 164 -
Each command must end with CR LF characters.
38.2. Response Format
On receipt of a command, the indicator responds as follows:
XX_A CR LF
command understood and in progress
XX_D CR LF
command carried out (appears only after the XX_A command)
XX_I CR LF
command understood but not accessible at this moment
XX _ ^ CR LF
command understood but max threshold is exceeded
XX _ v CR LF
command understood but min threshold is exceeded
XX _ OK CR LF
command carried out
ES_CR LF
command not recognised
XX _ E CR LF
time limit exceeded while waiting for stable measurement
result (time limit is a characteristic module parameter)
XX
- stands for a name of a sent command
_ - space
COMMANDS OVERVIEW
Zeroing
Format:
Z CR LF
Response options:
Z_A CR LF
- command understood and in progress
Z_D CR LF
- command carried out
Z_A CR LF
- command understood and in progress
Z_^ CR LF
- command understood but zeroing range is exceeded
Z_A CR LF
- command understood and in progress
Z_E CR LF
- time limit exceeded while waiting for stable measurement result
Z_I CR LF
- command understood but not accessible at this moment
Taring
Format:
T CR LF
Response options:
T_A CR LF
- command understood and in progress
T_D CR LF
- command carried out
T_A CR LF
- command understood and in progress
T_v CR LF
- command understood but taring range is exceeded
T_A CR LF
- command understood and in progress
T_E CR LF
- time limit exceeded while waiting for stable measurement result